Catalog Number 0606600000 |
Device Problem
Metal Shedding Debris (1804)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 05/01/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that after repair at the west distribution center (stryker (b)(4)) the revolutin cement gun was producing metal shavings.It was reported that there was no associated procedure.No patient involvement, no medical intervention and no adverse consequences were reported with this event.
|
|
Event Description
|
It was reported that after repair at the distribution center ((b)(4)) the revolution cement gun was producing metal shavings.It was reported that there was no associated procedure.No patient involvement, no medical intervention and no adverse consequences were reported with this event.
|
|
Manufacturer Narrative
|
A follow up report will be filed after the quality investigation has been completed.
|
|
Manufacturer Narrative
|
Upon visual inspection, the ram bushings were found to be in need of replacement.The device was repaired and returned to the user facility.
|
|