• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO CDI 500 BLOOD PARAMETER MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO CDI 500 BLOOD PARAMETER MONITOR Back to Search Results
Model Number 500AVHCT
Device Problem Low Test Results (2458)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/31/2014
Event Type  malfunction  
Event Description
It was reported that during use of the device for a cardiopulmonary bypass procedure, the customer encountered a low potassium (k+) value with the shunt sensor on the blood parameter monitor (bpm).The shunt sensor was changed out for a new one.However, the situation did not change.With the combination of the new shunt sensor and the same bpm unit, a low k+ value was obtained again.The customer gave up using the bpm unit that day.As a result, an alternate device was employed.They used cerebral oximetry.The surgical procedure was completed successfully.There were no delays, no blood loss, or no adverse consequences to the patient.
 
Manufacturer Narrative
Three days later on (b)(6) 2014, the bpm unit was used with another new shunt sensor in a procedure.There was no anomaly observed on the k+ value obtained from the combination of the bpm unit and the shunt sensor.The customer is not returning the unit at this time.They are still using the bpm unit for cases.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO CDI 500 BLOOD PARAMETER MONITOR
Type of Device
CDI 500
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson rd.
ann arbor, MI 48103
7346634145
MDR Report Key3832570
MDR Text Key4496700
Report Number1828100-2014-00235
Device Sequence Number1
Product Code DRY
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K972962
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 03/04/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number500AVHCT
Device Catalogue Number500AVHCT
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/04/2014
Initial Date FDA Received03/27/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
SHUNT SENSOR; CEREBRAL OXIMETRY
-
-