HOSPIRA COSTA RICA LTD. LC PCA NEW 802.11 AB; 80MEA Back to Search Results
Catalog Number 20709
Device Problems Occlusion Within Device (1423); Failure to Sense (1559)
Patient Problem No Patient Involvement (2645)
Event Date 02/18/2014
Event Type  malfunction  
Event Description
The customer contact reported that during preventive maintenance testing at the user facility, the device did not alarm when a distal occlusion was present.There were no reports of any adverse pt events and no reported delays of critical therapies while the device was in clinical use.Though requested, no additional information was provided.
 
Manufacturer Narrative
The device was received.Investigation is not complete.This report represents all the information known by the reporter upon query by hospira personnel.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LC PCA NEW 802.11 AB
Type of Device
80MEA
Manufacturer (Section D)
HOSPIRA COSTA RICA LTD.
zona franca global
la aurora heredia
CS 
Manufacturer (Section G)
HOSPIRA COSTA RICA LTD
zona franca global
la aurora heredia
CS  
Manufacturer Contact
juergen schmider, md, vice pres
275 n field drive
bldg no h2-1east dept no 097u
lake forest, IL 60045
2242125740
MDR Report Key3832826
MDR Text Key4501895
Report Number9615050-2014-02023
Device Sequence Number1
Product Code MEA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042800
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 02/19/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number20709
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/26/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/19/2014
Initial Date FDA Received03/17/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-