It was reported that during use of the device for a cardiopulmonary bypass (cpb) procedure, the hematocrit saturation module (h/sat) sensor cable was loose creating inaccurate h/sat reading.Example of reading: thirty-one reading was actually a twenty-two.The device was not changed out, as tension was applied on cable and the blood parameter monitor (bpm) corrected to appropriate value.They also used laboratory results to compare.The surgical procedure was completed successfully.There were no delays, no blood loss, or no adverse consequences to the patient.Per the clinical review on (b)(6) 2014: the blood parameter monitor (bpm) passed the color-chip test during set-up and boot-up of the device.During cpb, the h/sat data became erratic with values displaying dashes (---) or values being inaccurate.The inaccuracies were obvious and known to the user as the values changed dramatically, in spite of no dramatic clinical changes.An example ws the hematocrit was reading about twenty-two to twenty-three percent and closely matching the value of a laboratory analyzer and in just a few minutes with clinical changes, the hematocrit jumped to thirty-one percent.The perfusionist (ccp) felt the issue was with the cable (at junction to the probe head) as if the cable was flexed, reasonable values returned and would be displayed intermittently.The device was not changed out during the procedure and laboratory analyzer values were used to manage the patient (not the bpm values).The procedure was completed successfully, without delay and without associated blood loss (adult patient).There was no patient harm observed or reported.
|