|
Model Number H749518080 |
Device Problems
No Display/Image (1183); Difficult To Position (1467)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/25/2013 |
Event Type
malfunction
|
Event Description
|
Reportable based on analysis completed on (b)(4) 2014.It was reported that severe resistance and image lost were encountered.The opticross¿ imaging catheter was used to visualize a lesion it was noted that during percutaneous coronary intervention, severe resistance was encountered when imaging core was got back to the initial position and image was lost.The procedure was completed with another of the same device.No patient complications were reported and the patient's status is good.However, returned device analysis revealed an open hole in the sheath lap joint assembly.
|
|
Manufacturer Narrative
|
(b)(4).Device evaluated by mfr: the complaint device was received with a kink observed in the sheath assembly at 72.5 cm from femoral marker at the distal end.An open hole was observed in the sheath lap joint assembly 72.5 cm from the femoral marker at the distal.Fluid was leaking from an open hole in the lap joint assembly when the catheter was flushed.The telescope assembly was not able to properly pullback or advance.No wind up was observed on the imaging core during x-ray analysis.The ccp board pins appeared normal and a good click sound was heard during insertion into the mdu system.Full image characterization testing cannot be performed based on the returned condition of the catheter, however during impedance testing good image was observed.No other issues or defects were observed during product analysis of the returned device.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
|
|
Search Alerts/Recalls
|
|
|