• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOSPIRA COSTA RICA LTD. LIFECARE PCA 3 V5.06; 80MEA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOSPIRA COSTA RICA LTD. LIFECARE PCA 3 V5.06; 80MEA Back to Search Results
Catalog Number 20709
Device Problems Occlusion Within Device (1423); Failure to Sense (1559)
Patient Problem No Patient Involvement (2645)
Event Date 01/01/2014
Event Type  malfunction  
Event Description
The customer contact reported that during preventive maintenance testing at the user facility, the device did not alarm when a distal occlusion was present.There were no reports of any adverse patient events and no reported delays of critical therapies while the device was in clinical use.No additional information was provided.
 
Manufacturer Narrative
Investigation is not complete.This report represents all the information known by the reporter upon query by hospira personnel.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFECARE PCA 3 V5.06
Type of Device
80MEA
Manufacturer (Section D)
HOSPIRA COSTA RICA LTD.
zona franca global
la aurora heredia
CS 
Manufacturer (Section G)
HOSPIRA COSTA RICA LTD.
zona franca global
la aurora heredia
CS  
Manufacturer Contact
juergen schmider, md, vp
275 n. field dr.
bldg. no. h2-1east, dept no. 097u
lake forest, IL 60045
2242125740
MDR Report Key3837998
MDR Text Key15806439
Report Number9615050-2014-01828
Device Sequence Number1
Product Code MEA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042800
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Unknown
Reporter Occupation Other
Type of Report Initial
Report Date 02/12/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number20709
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/12/2014
Initial Date FDA Received03/10/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/01/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-