BOSTON SCIENTIFIC - FREMONT (SUD) OPTICROSS?; CATHETER, ULTRASOUND, INTRAVASCULAR Back to Search Results
Model Number H749518080
Device Problem Kinked (1339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/06/2014
Event Type  malfunction  
Event Description
Reportable based on device analysis completed on (b)(4) 2014.It was reported that during the unpacking of the opticross¿ imaging catheter, a kink was noted on the catheter.The procedure was completed with another of the same device.No patient complications were reported and the patient's condition is good.However, device analysis revealed a hole in the catheter.
 
Manufacturer Narrative
(b)(4).Device evaluated by manufacturer: the complaint device was received for evaluation.Evaluation of the returned device revealed that no kinks were observed along the length of the catheter, fluid was leaking from an open hole at the lap joint area when the catheter was flushed.The imaging window is still connected to the blue sheath at the lap joint.Leak is located on the outside edge of the device.Impedance testing finds the device is capable of imaging properly.Full image characterization testing cannot be performed due to the hole in the lap joint.It prevent properly flushing of the device.No other issues or defects were observed during product analysis of the returned device.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause was unable to be determined.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTICROSS?
Type of Device
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
BOSTON SCIENTIFIC - FREMONT (SUD)
47215 lakeview blvd phone
west dock
fremont CA 94538
Manufacturer (Section G)
BOSTON SCIENTIFIC - FREMONT (SUD)
47215 lakeview blvd phone
west dock
fremont CA 94538
Manufacturer Contact
ingrid matte
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key3838149
MDR Text Key4654919
Report Number2134265-2014-02931
Device Sequence Number1
Product Code OBJ
Combination Product (y/n)N
PMA/PMN Number
K123621
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 05/02/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/17/2014
Device Model NumberH749518080
Device Catalogue Number51808
Device Lot Number16466443
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/02/2014
Initial Date FDA Received05/29/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/17/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-