• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MIDMARK CORP PROGENY PREVA; UNIT, XRAY, EXTRAORAL WITH TIMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MIDMARK CORP PROGENY PREVA; UNIT, XRAY, EXTRAORAL WITH TIMER Back to Search Results
Model Number PREVA
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/26/2013
Event Type  malfunction  
Event Description
The field service tech of a distributor reported to midmark that during a field service to a preva x-ray unit, serial number (b)(4), he found that the unit generated a short x-ray exposure for approximately 200 milliseconds when it is turned on.The x-ray unit operated as intended after it was turned on.
 
Manufacturer Narrative
The field service engineer of the distributor performed an on-site investigation with remote support from midmark on (b)(6) 2014.The field service engineer reported that the failure was repeatable, and was eliminated after the control unit was replaced.The control unit was then returned to midmark for further investigation on (b)(6) 2014.However, the failure could not be replicated at midmark.The control unit passed all functional tests and visual inspection, and operated as intended at midmark.Midmark also performed schematic review and concluded that the reported failure mode cannot be caused by a single fault condition (i.E.Single component failure).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGENY PREVA
Type of Device
UNIT, XRAY, EXTRAORAL WITH TIMER
Manufacturer (Section D)
MIDMARK CORP
675 heathrow dr
lincolnshire IL 60069
Manufacturer Contact
peng lin
675 heathrow dr
lincolnshire, IL 60065
8474159787
MDR Report Key3843061
MDR Text Key4660193
Report Number1423380-2014-00001
Device Sequence Number1
Product Code EHD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043092
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Service Personnel
Type of Report Initial
Report Date 01/22/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberPREVA
Device Catalogue NumberP7016
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/10/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/26/2013
Initial Date FDA Received01/24/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/30/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-