• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BATTLE CREEK EQUIPMENT THERMOPHORE AUTOMATIC HEAT PACK; ELECTRIC MOST HEAT PACK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BATTLE CREEK EQUIPMENT THERMOPHORE AUTOMATIC HEAT PACK; ELECTRIC MOST HEAT PACK Back to Search Results
Model Number 255
Device Problem Fire (1245)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/30/2014
Event Type  malfunction  
Event Description
Consumer was laying on the pad in bed when the wire going to the switch caught fire.
 
Manufacturer Narrative
Our inspection found a small cut in the cord near the switch that could have exposed the user to bare wire.The inspection also revealed customer misuse as out inspection found: fda: cord cut/burned strain relief, switch ok, bent/broken lead, thermostat discoloration.Cord breakage unusually occurs when the cord is repeatedly over bent near the switch.We have initiated a capa project (b)(4) to reduce the occurrence of this issue.A new jacketed cord design was launched in production on (b)(6) 2014 that closes (b)(4) pending a future eval of the effectiveness of the solution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOPHORE AUTOMATIC HEAT PACK
Type of Device
ELECTRIC MOST HEAT PACK
Manufacturer (Section D)
BATTLE CREEK EQUIPMENT
fremont IN
Manufacturer Contact
randy newsome
702 south reed st.
fremont, IN 46737
2699626181
MDR Report Key3843820
MDR Text Key4409975
Report Number1811605-2014-00120
Device Sequence Number1
Product Code IRT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,User Facility
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 05/23/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number255
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/09/2014
Initial Date FDA Received05/24/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-