• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET ORTHOPEDICS JUGGERKNOT 1.5MM 1 #2 MB; FASTENER, FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET ORTHOPEDICS JUGGERKNOT 1.5MM 1 #2 MB; FASTENER, FIXATION Back to Search Results
Model Number N/A
Device Problem Device Slipped (1584)
Patient Problem No Information (3190)
Event Date 05/06/2014
Event Type  Injury  
Event Description
It was reported patient underwent an arthroscopic labral repair procedure on (b)(6) 2014.During the procedure, the suture anchor pulled out.An additional hole was drilled and another anchor was utilized to complete the procedure.
 
Manufacturer Narrative
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show that lot released with no recorded anomaly or deviation.There are warnings in the package insert that state that this type of event can occur: under warnings, number 2 states, "improper selection, placement, positioning, and fixation of the device can lead to failure of the device or the procedure.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JUGGERKNOT 1.5MM 1 #2 MB
Type of Device
FASTENER, FIXATION
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
amanda zajicek
56 e. bell drive
warsaw, IN 46582
5743726782
MDR Report Key3845646
MDR Text Key4616283
Report Number0001825034-2014-05163
Device Sequence Number1
Product Code MBI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK110145
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/08/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date03/31/2019
Device Model NumberN/A
Device Catalogue Number912031
Device Lot Number365830
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/08/2014
Initial Date FDA Received06/03/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/31/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-