• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: J. T. POSEY CO. POSEY KEEPSAFE DELUXE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

J. T. POSEY CO. POSEY KEEPSAFE DELUXE Back to Search Results
Model Number 8374
Device Problem Defective Alarm (1014)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
Customer reported the alarm does sound; however, the voice recording does not.Batteries have been replaced with a new supply and no visible damage reported.Customer could not provide the date when discovered.No pt incident or injury reported.
 
Manufacturer Narrative
Eval of the returned product confirmed the reported issue is confirmed.Factory message does not play and a clicking noise sounds in its place.Tone plays when it should.Unit passes all other functional tests.No physical damage observed.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POSEY KEEPSAFE DELUXE
Manufacturer (Section D)
J. T. POSEY CO.
arcadia CA
Manufacturer (Section G)
J. T. POSEY CO.
5635 peck rd.
arcadia CA 91006
Manufacturer Contact
pam wampler, admin
5635 peck rd.
arcadia, CA 91006
6264433143
MDR Report Key3847720
MDR Text Key4407622
Report Number2020362-2014-00077
Device Sequence Number1
Product Code KMI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 02/14/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8374
Device Catalogue Number8374
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/28/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/14/2014
Initial Date FDA Received03/18/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
OVER-THE-MATTRESS SENSOR PAD: MODEL 8307, LOT #UNK
-
-