• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION CHARGING SYSTEM; SCS CHARGING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION CHARGING SYSTEM; SCS CHARGING SYSTEM Back to Search Results
Model Number 3721
Device Problems Overheating of Device (1437); Defibrillation/Stimulation Problem (1573); Low Battery (2584)
Patient Problem No Information (3190)
Event Date 03/01/2014
Event Type  Injury  
Event Description
Device #2 of 2.Reference mfr.Report: 1627487-2014-26423.
 
Manufacturer Narrative
This charger model number was included in a field correction.Manufacturer's evaluation: corrective and preventive action (capa) investigation was performed.Evaluation: result: - pocket heating was confirmed.The investigation for capa (b)(4) associated with heating while charging (pocket heating) concluded that the charger was capable of transferring energy to the ipg at a rate that would cause heating of the ipg and/or charging wand of sufficient elevated temperature to cause pain and burns.The heating while charging was determined to be exacerbated by off-axis charging of shallow implanted ipgs and that all chargers were capable of elevated heating.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHARGING SYSTEM
Type of Device
SCS CHARGING SYSTEM
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano TX
Manufacturer Contact
tanja hoffman
6901 preston rd.
plano, TX 75024
9723098520
MDR Report Key3849079
MDR Text Key4681420
Report Number1627487-2014-26424
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Not Applicable
Remedial Action Other
Type of Report Initial
Report Date 04/29/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/01/2012
Device Model Number3721
Device Lot Number3575703
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/29/2014
Initial Date FDA Received05/20/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/01/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-07262012-001-C
Patient Sequence Number1
Treatment
IMPLANT DATE:; SCS LEAD, MODEL: 3228
Patient Outcome(s) Other;
Patient Age62 YR
-
-