The actual device was returned to the richard wolf facility and investigation completed (b)(4) 2014.Upon visual and functional inspection by the optics department, the adhesive around the distal hose was missing.The root cause is reprocessing.Adhesive was re-applied and tested prior to returned to customer.Labeling was reviewed and found to be adequate.Ie intended use, indications and field of use, preparation and cautions.Rwmic considers this matter closed; however, in the event additional information is received, we will provide the fda with follow-up information.
|