• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - IRVINE INFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - IRVINE INFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C1009KP
Device Problems Excess Flow or Over-Infusion (1311); Positioning Problem (3009)
Patient Problem Overdose (1988)
Event Date 04/30/2014
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This lot was manufactured between november 26, 2013 and november 27.Evaluation summary: baxter received one unit for evaluation.Visual inspection showed no signs of physical abnormality.A functional flow rate test was performed on the unit and was found to be within specification.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a large volume infusor overinfused.The reporter stated that the device was filled with fluorouracil and was expected to fully deliver its contents in 46 hours, but the infusion was completed in 34 hours.The patient stated that they were not in a shower or heated area, the device was around their neck, and they slept on their side all night.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
(b)(4).The device was returned and is currently in the process of being evaluated.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
bldg 3
irvine CA 92614
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
bldg 3
irvine CA 92614
Manufacturer Contact
christina arnt
25212 w. illinois route 120
round lake, IL 60073
2242703198
MDR Report Key3851071
MDR Text Key16816480
Report Number1416980-2014-17976
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K071222
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 05/12/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/01/2016
Device Catalogue Number2C1009KP
Device Lot Number13M091
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/30/2014
Initial Date FDA Received06/05/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/24/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/27/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
FLUOROURACIL
-
-