MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD TEMPERATURE/FLOW PROBE; BTT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD TEMPERATURE/FLOW PROBE; BTT Back to Search Results
Model Number 900MR869
Device Problem Disconnection (1171)
Patient Problem Low Oxygen Saturation (2477)
Event Type  malfunction  
Event Description
A hospital in the (b)(6) reported to a fisher & paykel healthcare (fph) representative that a 900mr869 temperature/flow probe became disconnected from the rt330 infant optiflow circuit during use.They further reported that the patient began to desaturate.
 
Manufacturer Narrative
(b)(4).Our investigation is currently in progress.We will provide a follow up report upon completion of our investigation.
 
Manufacturer Narrative
(b)(4).Method: the complaint 900mr869 temperature probe was not returned to fisher & paykel healthcare in (b)(4) for evaluation.Therefore, our investigation is based on the information provided by the hospital, information gathered by an fph representative during a visit to the hospital, and our knowledge of the product.Results: during the fph rep's visit it was noticed that the hospital staff were not completely pushing in the 900mr869 temperature probe.Conclusion: without the return of the complaint device we are unable to determine the root cause of the problem reported by the customer.However, the reported disconnection was most likely due to the temperature probe not properly inserted into the inspiratory limb of the rt330 infant optiflow circuit.Since the reported incident a fisher & paykel healthcare representative has visited the hospital and has provided additional training on the correct placement of the temperature probes.We have had no further reports from this hospital.The user instructions that accompany the 900mr869 temperature probe state: "the probes must also be pushed firmly into the probe ports on the inspiratory side of the breathing circuit and the key on the chamber outlet probe must align with the probe port key-way.Furthermore, the user instructions that accompany the rt330 infant optiflow circuit state: "check that all connections, caps and/or plugs are tight before use." "patient monitoring is recommended.".
 
Event Description
A hospital in the (b)(6) reported to a fisher & paykel healthcare (fph) representative that a 900mr869 temperature/flow probe became disconnected from the rt330 infant optiflow circuit during use.They further reported that the patient began to desaturate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPERATURE/FLOW PROBE
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
raymond yan
15365 barranca parkway
irvine, CA 92618-2216
9494534000
MDR Report Key3852448
MDR Text Key4681495
Report Number9611451-2014-00491
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K892885
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 05/07/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number900MR869
Device Catalogue Number900MR869
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/07/2014
Initial Date FDA Received06/05/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/14/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
FPH RT330 INFANT OPTIFLOW CIRCUIT; FPH MR850 RESPIRATORY HUMIDIFIER; FPH RT330 INFANT OPTIFLOW CIRCUIT; FPH MR850 RESPIRATORY HUMIDIFIER
-
-