Catalog Number 136506000
Device Problems Loss of Osseointegration (2408); Naturally Worn (2988)
Patient Problems Host-Tissue Reaction (1297); Ossification (1428); Adhesion(s) (1695); Cyst(s) (1800); Foreign Body Reaction (1868); Unspecified Infection (1930); Muscular Rigidity (1968); Scar Tissue (2060); Discomfort (2330); No Information (3190); No Code Available (3191); Muscle/Tendon Damage (4532); Unspecified Musculoskeletal problem (4535); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 10/13/2011
Event Type  Injury  
Event Description
Litigation alleges loosening.
 
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
 
Manufacturer Narrative
The device associated with this report was not returned.A complaint database search finds no other reported incidents against the provided product and lot combination.The investigation could not verify or identify any product contribution to the reported event with the information provided.Based on the inability to identify root cause, the need for corrective action was not indicated.Depuy considers the investigation closed at this time.Should the product and/or additional information be received, the investigation will be re-opened.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Udi: (b)(4).
 
Event Description
Ppf alleges infection, metal wear, metallosis and elevated metal ions.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.  h10 additional narrative:   udi: (b)(4).Corrected: h6 (patient).
 
Event Description
Updated code for metallosis.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: investigation summary
=
> no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.No device associated with this report was received for examination.The adverse symptoms alleged and product code reported is associated with the depuy metal on metal articulation.A complaint database search and/or device manufacturing (dhr) reviews will not be performed.Per wi-3430 it has been determined that should related reports be identified a dhr review is not required.Investigational inputs were requested as indicated per internal procedures for this failure mode.The complaint information provided has been reviewed for complaint coding, medical device reporting, and other data required by the complaint system.Per internal procedures, the event information and any investigational inputs received as part of required follow up were reviewed.For this investigation, no immediate action was required as no alleged deficiency with the device(s) was identified.- without the physical complaint sample(s) associated with this report, it was not possible to determine if the device(s) failed to meet specification(s) at the time it was released for distribution.- the device(s) associated with this event were used in the treatment of the patient as prescribed by the presiding surgeon.- from the event information received, it was not possible to determine the relationship of the device to the reported event.No information received with this individual complaint indicated that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H6 clinical code: appropriate term / code not available (e2402) is used to capture blood heavy metal increased.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
In addition to what previously alleged, pfs alleges stiffness and discomfort.After review of medical records patient was revised was due to loose acetabular component.Operative notes indicated that there were osteophytes present superiorly and anteriorly as well as soft tissue had to be debrided.The broken screw could be seen at that point.It appeared that the flange on the screw head had broken off or worn away and the screw had bent with almost 90 degrees angle posteriorly as the cup traveled posteriorly.The overall quality of the acetabulum was good.There were cystic areas but there were also very sclerotic areas an unusual undulating areas of sclerotic bone presumably due to metal impingement and reaction to wear.There were adhesion present, tendonitis, some slightly grayish fluid and the screw was sitting proud.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : a manufacturing record evaluation (mre) will not be performed since mom systems are obsolete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12/14 ARTICUL 40MM M SPEC+5
Type of Device
HIP FEMORAL HEAD
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
LEEDS MFG & MATERIAL WAREHOUSE
st anthonys road
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kara ditty-bovard
700 orthopaedic drive
warsaw, IN 46581-0988
6107428552
MDR Report Key3864121
MDR Text Key4509616
Report Number1818910-2014-20594
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK060031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Attorney
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 06/18/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date04/30/2013
Device Catalogue Number136506000
Device Lot Number2590333
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/18/2014
Initial Date FDA Received06/10/2014
Supplement Dates Manufacturer ReceivedNot provided
04/02/2019
09/26/2019
10/17/2019
11/16/2021
02/10/2022
Supplement Dates FDA Received06/30/2014
04/19/2019
10/14/2019
11/06/2019
11/30/2021
02/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/12/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
12/14 ARTICUL 40MM M SPEC+5; PINN CAN BONE SCREW 6.5MMX30MM; PINN SECTOR W/GRIPTION 56MM; PINNACLE MTL INS NEUT40IDX56OD; SUMMIT POR TAPER SZ6 HI OFF; UNKNOWN HIP IMPLANT
Patient Outcome(s) Required Intervention;
Patient Age44 YR
Patient SexMale
Patient Weight114 KG
-
-