Catalog Number 136506000 |
Device Problems
Loss of Osseointegration (2408); Naturally Worn (2988)
|
Patient Problems
Host-Tissue Reaction (1297); Ossification (1428); Adhesion(s) (1695); Cyst(s) (1800); Foreign Body Reaction (1868); Unspecified Infection (1930); Muscular Rigidity (1968); Scar Tissue (2060); Discomfort (2330); No Information (3190); No Code Available (3191); Muscle/Tendon Damage (4532); Unspecified Musculoskeletal problem (4535); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 10/13/2011 |
Event Type
Injury
|
Event Description
|
Litigation alleges loosening.
|
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
|
|
Manufacturer Narrative
|
The device associated with this report was not returned.A complaint database search finds no other reported incidents against the provided product and lot combination.The investigation could not verify or identify any product contribution to the reported event with the information provided.Based on the inability to identify root cause, the need for corrective action was not indicated.Depuy considers the investigation closed at this time.Should the product and/or additional information be received, the investigation will be re-opened.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Udi: (b)(4).
|
|
Event Description
|
Ppf alleges infection, metal wear, metallosis and elevated metal ions.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report. h10 additional narrative: udi: (b)(4).Corrected: h6 (patient).
|
|
Event Description
|
Updated code for metallosis.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: investigation summary = > no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.No device associated with this report was received for examination.The adverse symptoms alleged and product code reported is associated with the depuy metal on metal articulation.A complaint database search and/or device manufacturing (dhr) reviews will not be performed.Per wi-3430 it has been determined that should related reports be identified a dhr review is not required.Investigational inputs were requested as indicated per internal procedures for this failure mode.The complaint information provided has been reviewed for complaint coding, medical device reporting, and other data required by the complaint system.Per internal procedures, the event information and any investigational inputs received as part of required follow up were reviewed.For this investigation, no immediate action was required as no alleged deficiency with the device(s) was identified.- without the physical complaint sample(s) associated with this report, it was not possible to determine if the device(s) failed to meet specification(s) at the time it was released for distribution.- the device(s) associated with this event were used in the treatment of the patient as prescribed by the presiding surgeon.- from the event information received, it was not possible to determine the relationship of the device to the reported event.No information received with this individual complaint indicated that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H6 clinical code: appropriate term / code not available (e2402) is used to capture blood heavy metal increased.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
In addition to what previously alleged, pfs alleges stiffness and discomfort.After review of medical records patient was revised was due to loose acetabular component.Operative notes indicated that there were osteophytes present superiorly and anteriorly as well as soft tissue had to be debrided.The broken screw could be seen at that point.It appeared that the flange on the screw head had broken off or worn away and the screw had bent with almost 90 degrees angle posteriorly as the cup traveled posteriorly.The overall quality of the acetabulum was good.There were cystic areas but there were also very sclerotic areas an unusual undulating areas of sclerotic bone presumably due to metal impingement and reaction to wear.There were adhesion present, tendonitis, some slightly grayish fluid and the screw was sitting proud.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : a manufacturing record evaluation (mre) will not be performed since mom systems are obsolete.
|
|