Asp investigation summary: the investigation included a review of the device history record (dhr), service history, trending of the product malfunction code, and failure mode and effects analysis (fmea).The dhr was not reviewed as the lot number of the cassette was not available.The trend for the product malfunction code of skin reaction was assessed from june 2013 through may 2014.No significant trend was found.The fmea revealed the risk priority number (rpn) scores are considered to be acceptable.Per the 100nx cassette instructions for use: observe the leak indicator strip on the wrapped cassette.If the leak indicator strip indicates ¿no leak¿, it is ok to handle the cassette.If the indicator strip indicates a ¿leak¿, do not handle the cassette.No product was returned for further evaluation.Review of the tracking and trending data did not identify a trend.As a result, root cause or assignable cause analysis could not be performed.No further action is required; however, this issue will continue to be monitored.
|