• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO PERFUSION SYSTEM 8000; 8K (PUMP, ROLLER TYPE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO PERFUSION SYSTEM 8000; 8K (PUMP, ROLLER TYPE) Back to Search Results
Model Number 16402
Device Problems Device Operates Differently Than Expected (2913); Positioning Problem (3009)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/01/2014
Event Type  malfunction  
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the roller pump occlusion mechanism was stuck in the open position.Perfusionist (ccp) was not able to turn occlusion knob.As a result, an alternate device was employed.The surgical procedure was completed successfully.There were no delays, no blood loss, or no adverse consequences to the patient.
 
Manufacturer Narrative
Evaluation is in progress, but not yet concluded.
 
Manufacturer Narrative
Updated block.The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Manufacturer Narrative
Per the laboratory evaluation, the technician observed the pump occlusion was stuck upon receipt of the pump.The occlusion was able to be loosened after removal of the occluder knob and tongue, and functioned after the occluder knob was placed back on the pump.Abnormal wear was seen on the screw adjust slot where the tongue fits.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO PERFUSION SYSTEM 8000
Type of Device
8K (PUMP, ROLLER TYPE)
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson rd.
ann arbor, MI 48103
7346634145
MDR Report Key3869921
MDR Text Key4591337
Report Number1828100-2014-00288
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K953901
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,health professional,use
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 04/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16402
Device Catalogue Number16402
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer04/21/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/01/2014
Initial Date FDA Received04/25/2014
Supplement Dates Manufacturer Received04/15/2016
04/15/2016
Supplement Dates FDA Received06/23/2014
04/18/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-