U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND SENSOR MODULE LEVEL/BUBBLE; MONITOR AND/OR CONTROL, LEVEL SENSING, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND SENSOR MODULE LEVEL/BUBBLE; MONITOR AND/OR CONTROL, LEVEL SENSING, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 23-30-20
Device Problem Other (for use when an appropriate device code cannot be identified) (2203)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/08/2014
Event Type  malfunction  
Event Description
Sorin group (b)(4) received a report that the sensor module level/bubble alarmed and the pump stopped during a procedure.There was no report of patient injury.
 
Manufacturer Narrative
Sorin group (b)(4) manufactures the sensor module level/bubble.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the sensor module level/bubble alarmed and the pump stopped during a procedure.There was no report of patient injury.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Manufacturer Narrative
Sorin group (b)(6) manufactures the sensor module level/bubble.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group received a report that the sensor module level/bubble alarmed and the pump stopped during a procedure.There was no report of patient injury.A sorin group representative went on site and was unable to reproduce the reported issue.The bubble sensor and arterial pump along with all connections were checked and no issues were discovered.Loan modules were installed and the replaced level/bubble display module, level/bubble sensor module and dc/dc module were sent to sorin group (b)(4) for further evaluation.The investigation performed by sorin group (b)(4) was also unable to duplicate the issue.A test run of 24 hours with various adjustments and different rotational speeds was performed and no deviations were found during this time.The system worked as specified and no report of recurrence has been received.A review of the dhr was unable to identify and concessions, deviations or non-conformities relevant to the reported issue.Sorin group (b)(4) will continue to monitor the market for trends related to this issue and if any are identified, corrections will be recommended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR MODULE LEVEL/BUBBLE
Type of Device
MONITOR AND/OR CONTROL, LEVEL SENSING, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich D 809 39
GM  D 80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
carrie wood
14401 w. 65th way
arvada, CO 80004
3034676461
MDR Report Key3878048
MDR Text Key4595049
Report Number1718850-2014-00165
Device Sequence Number1
Product Code DTW
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K955152
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,health professional
Reporter Occupation Unknown
Remedial Action Other
Type of Report Initial,Followup
Report Date 05/09/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number23-30-20
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 05/09/2014
Initial Date FDA Received06/05/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/23/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2001
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age89 YR
-
-