Device Problem Use of Device Problem (1670)
Patient Problem Foreign Body In Patient (2687)
Event Type  Injury  
Event Description
Report received of a user error resulting in a suction swab disengagement.Reporter stated that on (b)(6) 2014, pt had a scheduled esophagogastroduodenoscopy (egd) the reason for procedure was unk by reporter.During the esg, an object described as a portion of green foam with a small white plastic tube in the center was noted in the pt's stomach.Reporter identified the object as the tip of a sage suction swab.Reporter stated the object was removed so that it would not obstruct the drainage flow of pt's nasogastric tube.Reporter stated upon viewing object, it was evident the pt had bitten off the tip of the suction swab.Reporter described involved pt as confused, disoriented, uncooperative with a history of biting.Reporter indicated that while hospitalized, there was no report of the pt biting off a suction swab.Reporter stated she was certain incident occurred in pt's home prior to hospitalization although the family never mentioned that fact, when asked about the lot number, expiration date, and scu number, as reported on form 3500, reporter stated she did not know the details from what kit the swab came.Reporter explained she selected a sage brand #(b)(4) q4 complete care kit from the storeroom, as it contained suction swabs, and completed section e of the form 3500 using that kit's lot number, expiration date, and scu.She did not inquire of the family about the brand of product used at home.Through requested, no additional info was available.
 
Manufacturer Narrative
Reporter declined request to return object for eval, although provided two photographs.Though reporter identified object as a sage suction swab, no model number, catalog number, or lot info was available.Eval of the photographs indicates the reported event is not attributable to a product defect and/or manufacturing operations.Eval of the photographs is consistent with reporter's statement that the object, which appears to be the tip of a suction swab, was bitten off by the pt.The packaging label contains precautionary statements advising the end user to use a bite block when performing oral care on pts with altered levels of consciousness, or those who cannot comprehend commands, and ensuring the foam head is intact.User facility report form fda 3500 received and reviewed.Provided on this form is additional, corrected, and missing info related to the reported event.Info was obtained after speaking with initial reporter.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
SUCTION SWAB
Manufacturer (Section D)
SAGE PRODUCTS, LLC
cary IL
Manufacturer Contact
denise hummel, rn, bsn
3909 three oaks rd
cary, IL 60013
8003232220
MDR Report Key3880655
MDR Text Key4458866
Report Number1419181-2014-00001
Device Sequence Number1
Product Code MMK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 05/02/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/02/2014
Initial Date FDA Received05/29/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
ENDOTRACHIAL TUBE; NASOGASTRIC TUBE
Patient Outcome(s) Required Intervention;
Patient Age61 YR
-
-