Following the reported event, the facility contacted steris service as they requested the surgical table be inspected for proper operation.A steris service technician arrived at the facility, inspected the table and found the equipment operating to specification.No faults with the table were identified and the table was turned back to the facility for use.The technician was informed by facility staff that the event occurred as the staff was moving the patient off the table to a stretcher for transport to another location.The steris account manager contacted the facility to provide additional in-service training on the proper use and operation of the surgical table however, the user facility was unresponsive.The table was installed in december of 2012 and is currently under steris extended warranty.No further issues have been reported.
|