|
Model Number H749518080 |
Device Problems
No Display/Image (1183); Kinked (1339)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/06/2014 |
Event Type
malfunction
|
Event Description
|
Reportable based on device analysis completed on (b)(4) 2014.It was reported that during a percutaneous coronary intervention (pci), kink and lost image occurred.The target lesion was located in the 100% stenosed, moderately tortuous and mildly calcified right coronary artery.The physician tried to repositioned the telescope of the opticross¿ imaging catheter after pullback however, the catheter got kinked and image disappeared.The procedure was completed with another of the same device.No patient complications were reported and the patient's condition is good.However, device analysis revealed a hole in the sheath lap joint of the catheter.
|
|
Manufacturer Narrative
|
Age at time of event: 18 years or older.(b)(4).Device evaluated by manufacturer: the complaint device was received for evaluation.Evaluation of the returned device revealed that a kink was observed in the imaging window assembly at 67.1 cm from femoral marker at the distal end.An open hole was observed at the sheath lap joint section of the device.Fluid was leaking from the open hole at the sheath lap joint assembly when the catheter was flushed.The telescope assembly was not able to properly pull back, advance, or retract.The telescope cannot advance the transducer distal housing (tdh) to the most distal position.During image characterization testing, no image appeared in the system due to electrical open at proximal.Imaging core windup was found within the telescope section of the device.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|
|
Search Alerts/Recalls
|
|
|