IRIS INTERNATIONAL ICHEM VELOCITY; AUTOMATED URINE CHEMISTRY ANALYZER Back to Search Results
Catalog Number 700-7177-001
Device Problems Communication or Transmission Problem (2896); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/02/2014
Event Type  malfunction  
Event Description
It was reported barcode was mis-reading samples.
 
Manufacturer Narrative
It was reported a sample was run and released twice, with the same id#, to lis.The result was held in lis due to flag that result for that id# was already done.The names did not match and the actual sample id# on the tube was different than what was read/sent to lis.No injuries or change to patient management reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICHEM VELOCITY
Type of Device
AUTOMATED URINE CHEMISTRY ANALYZER
Manufacturer (Section D)
IRIS INTERNATIONAL
9172 eton ave.
chatsworth CA 91311
Manufacturer Contact
sudha gupta
9172 eton ave.
chatsworth, CA 91311
8185277272
MDR Report Key3897456
MDR Text Key4425161
Report Number2023446-2014-00060
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101852
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/02/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number700-7177-001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/02/2014
Initial Date FDA Received05/22/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-