Customer reported the thinprep 5000 processor produced an error (se 6006 sh 024), also broke some slides (before putting cells on them), and lost some sample (non-gyn, urocyte).The patient will be called back.The customer stopped the unit.Hologic field service engineer confirmed and reproduced error (se 6006 sh 024).Found the following to be the most likely cause of the error: dirty vacuum cups (cleaned).Replaced 3v battery cr 2032 (sbc).Ran all non-interactive motion and go no-go tests to verify instrument operation.Processed samples to confirm operation.Instrument operational.Spain: this is not classified as a reportable event since this is a technical issue.The instrument issued a warning/error code, which alerted the user.The dispatched engineer was able to confirm and reproduce that error and resolved the issue by technical/service intervention.
|