Catalog Number 625-0T-28D |
Device Problem
Device Dislodged or Dislocated (2923)
|
Patient Problem
Toxicity (2333)
|
Event Date 05/07/2014 |
Event Type
Injury
|
Event Description
|
It was reported that as per the surgeon there was impingement at the 2 o'clock position with evidence of cocr metallosis.Previous dictation shows 1 dislocation immediately post op of original surgery with recurrent subluxation.
|
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of the investigation.
|
|
Manufacturer Narrative
|
An event regarding dislocation involving a trident liner was reported.The event was confirmed.A material analysis indicated the damage on the returned liner may have generated metallic debris in vivo.This is as a result of impingement due to a malpositioned cup.Medical records indicate - patient-related factors of high activity level associated with young age plus early postoperative capsular laxity potentially might have acted as secondary factor although cup position remains the principal problem.As such root causer of failure is mainly procedure-related with possible a secondary contribution by patient-related factors without evidence for device-related factors as also supported by the mar findings.This pi case is not device-related.A device history review indicated that the device was manufactured and accepted into final stock.There have been no other events for the reported lot.The results of the investigation indicate the dislocation, impingement and metal debris were caused by malposition of the cup and a highly active patient.The reported event is not device related.
|
|
Event Description
|
It was reported that as per the surgeon there was impingement at the 2 o'clock position with evidence of cocr metallosis.Previous dictation shows 1 dislocation immediately post op of original surgery with recurrent subluxation.
|
|
Search Alerts/Recalls
|