A discordant, falsely elevated chloride (cl) result was obtained on one patient sample on an advia 2400 instrument.The discordant result was released to the pathologist(s), who questioned it.The sample was repeated on an alternate instrument, resulting lower.The corrected result was reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated cl result.
|
The customer contacted the siemens technical solutions center (tsc).The customer stated that after replacing the ion selective electrode (ise) buffer, they did not perform ise calibrations prior to running patient samples.The customer re-calibrated the ise and ran quality controls, resulting within range.The cause of the discordant, falsely elevated cl result is a user error.The instrument is performing according to specifications.No further evaluation of the device is required.
|