STRYKER INSTRUMENTS-KALAMAZOO FAN SPRAY TIP WITH SOFT CONE SPLASH SHIELD; LAVAGE, JET Back to Search Results
Catalog Number 0207055000
Device Problem Disassembly (1168)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/03/2014
Event Type  malfunction  
Event Description
It was reported that the fan spray tip with soft cone splash shield was being used in a procedure when the tip broke off and came out of the gun.It was confirmed that the tip did not fall into the surgical site.The procedure was completed successfully, with no patient or user injuries and no adverse consequences.
 
Event Description
It was reported that the fan spray tip with soft cone splash shield was being used in a procedure when the tip broke off and came out of the gun.It was confirmed that the tip did not fall into the surgical site.The procedure was completed successfully, with no patient or user injuries and no adverse consequences.
 
Event Description
It was reported that the fan spray tip with soft cone splash shield was being used in a procedure when the tip broke off and came out of the gun.It was confirmed that the tip did not fall into the surgical site.The procedure was completed successfully, with no patient or user injuries and no adverse consequences.
 
Manufacturer Narrative
Upon visual inspection of the tip, the claimed condition was confirmed.The suction tube was detached/dislodged from the tip insert, and the irrigation tube was broken (possibly as result of the detachment of the suction tube).Glue residues were observed in the end of the suction tube, where it connects with the tip insert.At this time a specific root cause cannot be determined.The device was scrapped at the manufacturer.
 
Manufacturer Narrative
A follow up report will be filed after the device is received and the quality investigation has been completed.
 
Manufacturer Narrative
The surgilav plus handpiece set, (b)(4), lot no 14049012, was investigated and deemed a concomitant product to this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAN SPRAY TIP WITH SOFT CONE SPLASH SHIELD
Type of Device
LAVAGE, JET
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
highway #3, km 130.2
kalamazoo MI 49001 006
Manufacturer (Section G)
STRYKER INSTRUMENTS-PUERTO RICO
las palmas industrial park
highway #3, km 130.2
arroyo MI 49001 006
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key3900689
MDR Text Key4428554
Report Number0001811755-2014-02308
Device Sequence Number1
Product Code FQH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 06/03/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0207055000
Device Lot Number14111012
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/29/2014
Initial Date FDA Received06/27/2014
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received08/13/2014
08/15/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SURGILAV PLUS HP SET (B)(4), LOT: 04049012
-
-