BAXTER HEALTHCARE - IRVINE FOLFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C4702K
Device Problem Short Fill (1575)
Patient Problem Underdose (2542)
Event Date 06/04/2014
Event Type  malfunction  
Event Description
It was reported that a small volume folfusor underinfused.This occurred during infusion of 5 g/100 ml of fluorouracil.The device was filled with 96 ml of solution and was expected to infuse for 48 hours.The reporter stated that about 75% of the fill volume remained at the end of the infusion.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
(b)(4).(b)(6).The lot 14a061 was manufactured between january 30, 2014 to january 31, 2014.A batch review was conducted and revealed that all of the acceptance criteria were met to release the lot.Evaluation summary: the actual device was evaluated.Visual inspection of the device did not reveal any abnormalities that could have contributed to the reported condition.A functional flow rate testing was performed and the device was found within specification.The cause of the condition could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
bldg 3
irvine CA 92614
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
bldg 3
irvine CA 92614
Manufacturer Contact
christina arnt
25212 w. illinois route 120
round lake, IL 60073
2242703198
MDR Report Key3901408
MDR Text Key18064690
Report Number1416980-2014-20778
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 06/04/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/01/2016
Device Catalogue Number2C4702K
Device Lot Number14A061
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/04/2014
Initial Date FDA Received06/30/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
FLUOROURACIL (PFIZER)
Patient Age51 YR
Patient Weight80
-
-