Catalog Number 0206600000 |
Device Problem
Disassembly (1168)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 02/20/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that during testing conducted at the manufacturer facility that the head of screw on the dual speed cement injection gun was found to be missing.It was reported that there was no associated procedure.No patient involvement, no delay, no medical intervention and no adverse consequences were reported with this event.
|
|
Manufacturer Narrative
|
A follow up report will be filed after the quality investigation has been completed.
|
|
Event Description
|
It was reported that during testing conducted at the manufacturer facility that the head of screw on the dual speed cement injection gun was found to be missing.It was reported that there was no associated procedure.No patient involvement, no delay, no medical intervention and no adverse consequences were reported with this event.
|
|
Manufacturer Narrative
|
During the device evaluation, one of the cartridge interface screws was observed to be broken off, and corrosion was observed at the ram bushings.The device was repaired and returned to the user facility.
|
|