It was reported that device diagnostics resulted in high impedance (5348 ohms).The generator was programmed off and x-rays were taken.It was reported that the patient has been digging at the incision site since implant.It was reported that the patient was recently implanted.The patient was seen by the physician again and device diagnostics were performed which resulted in impedance values from 5200 ohms to high in the 5000 range.The physician indicated that the x-rays were ok.X-rays were sent to manufacturer for review.Based on the x-rays received, it appears that the lead pin not being fully inserted may be the cause of the device¿s high impedance.There was nothing seen that would indicate there was any damage to the generator or lead; however, the presence of a micro-fracture in the lead cannot be ruled out.As the entire lead could not be assessed, continuity in that portion of the lead cannot be confirmed.Surgery is likely, but has not been performed to date.
|