Catalog Number 2C1009KP
Device Problems Device Slipped (1584); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/11/2014
Event Type  malfunction  
Event Description
It was reported that there was particulate matter in a large volume infusor.This was noted after filling with fluorouracil.The reporter described the particulate matter as a loose fiber that was ¿attached to the filter on the stress member.¿ there was no patient involvement.No additional information is available.
 
Manufacturer Narrative
(b)(4).(b)(6).The device has been received and the evaluation is in progress.Upon completion of the investigation, or if any additional relevant information is received, a supplemental report will be submitted.
 
Manufacturer Narrative
(b)(4).The device was manufactured february 8, 2014 ¿ february 10, 2014.Evaluation summary: one sample was received for evaluation.Visual inspection noted a white thread-like particle 4.5 mm in length hanging from the stress-member filter.The particle was identified to be a polyester material via fourier transform infrared (ft-ir) spectrophotometer scanning.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The cause of the particulate matter was due to the storage and handling of the stress member subassemblies.There is a capa open to further investigate this issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
bldg 3
irvine CA 92614
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
bldg 3
irvine CA 92614
Manufacturer Contact
christina arnt
25212 w. illinois route 120
round lake, IL 60073
2242703198
MDR Report Key3910946
MDR Text Key4587536
Report Number1416980-2014-21369
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeEI
PMA/PMN Number
K071222
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/11/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date02/01/2017
Device Catalogue Number2C1009KP
Device Lot Number14B027
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/09/2014
Initial Date FDA Received07/02/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/01/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-