• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS RAPIDPOINT 500; RP 500

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS RAPIDPOINT 500; RP 500 Back to Search Results
Catalog Number 10492730
Device Problem Use of Incorrect Control/Treatment Settings (1126)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/07/2014
Event Type  malfunction  
Event Description
Customer reported that automatic quality control(aqc) was deselected in the setup on the instrument during (b)(6) 2014.Customer also reported that patient samples were run during this timeframe.There was no report of injury due to this event.
 
Manufacturer Narrative
Customer should not have run patient samples when automatic quality control (aqc) was turned off on the instrument.Customer confirmed that no patient results were questioned during this time frame and no treatment was affected.Siemens has issued siemens knowledge base (skb) in order to inform customers to turn automatic quality control (aqc) back on before running patient samples on the instrument.The event has occurred due to an operator error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDPOINT 500
Type of Device
RP 500
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industry
sudbury CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key3912367
MDR Text Key4663344
Report Number1217157-2014-00097
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/09/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Health Care Professional
Device Catalogue Number10492730
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/09/2014
Initial Date FDA Received07/03/2014
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-