• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC DIMENSION VISTA 1500; CLINICAL CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC DIMENSION VISTA 1500; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Model Number DIMENSION VISTA 1500
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/27/2014
Event Type  malfunction  
Event Description
A discordant, falsely elevated chloride (cl) result was obtained on one patient sample on a dimension vista 1500 instrument.The discordant result was not reported to the physician(s).The sample was repeated twice on the same instrument, resulting lower.It is unknown if the corrected result was reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated cl result.
 
Manufacturer Narrative
A siemens customer service engineer (cse) was dispatched to the customer site.After evaluation of the instrument and instrument data, the cse cleaned the system, replaced the pump tubing, replaced and aligned the integrated multisensor technology probe and aliquot probe and aligned the imt rotary valve.The cause of the discordant, falsely elevated cl result is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIMENSION VISTA 1500
Type of Device
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC
101 silvermine road
brookfield CT 06804
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS INC
101 silvermine road
brookfield CT 06804
Manufacturer Contact
cassandra kocsis
511 benedict ave
tarrytown, NY 10591
9145242687
MDR Report Key3913457
MDR Text Key20657473
Report Number1226181-2014-00346
Device Sequence Number1
Product Code CGZ
Combination Product (y/n)N
Reporter Country CodeDA
PMA/PMN Number
K051087
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/11/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDIMENSION VISTA 1500
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/11/2014
Initial Date FDA Received07/03/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/27/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-