Model Number MODEL 250
Device Problem Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/05/2014
Event Type  malfunction  
Event Description
It was reported that the physician's handheld was stuck at the align screen and he can't get it to go past that.A hard reset was performed several times; however, was not successful in getting the screen past the align screen.The physician was provided with a new programming tablet and the handheld was received for analysis.Analysis is underway, but has not been completed to date.
 
Manufacturer Narrative
 
Event Description
Analysis of the returned handheld device and software flashcard was completed.No anomalies associated with the handheld performance were noted during testing using the ac adapter or the main battery with a full charge.The handheld performed according to functional specifications.No anomalies associated with flashcard software or databases were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING SOFTWARE
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058 770
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key3913948
MDR Text Key4493007
Report Number1644487-2014-01685
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/05/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1073772
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/20/2014
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 07/10/2014
Initial Date FDA Received07/03/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/04/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/09/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-