Boston scientific received information that this implantable cardioverter defibrillator (icd) and right ventricular (rv) lead displayed a gradual increase in shock lead impedance (sli) measurements from 39 ohms to 102 ohms.Also, extended charge times were observed however remain within the extended charge time limit.The source of this clinical observations was not determined.A 1.1 joule synchronized shock and full energy shock was delivered which resulted in 87 ohms and 97 ohms, respectively.Two defibrillation threshold (dfts) tests were also performed with shocking impedance measurements of 98 ohms and 89 ohms.The patient will continue to be monitored for any out-of-range (oor) shocking impedance values.Additional information was obtained indicating that the patient heard the device was beeping.The field representative said that the device had not reached elective replacement indicator (eri) but the sli measurements were now up around 124 ohms.All other measurements were fine.They verified that the cause of beeping was an oor message in the daily test.Boston scientific technical services (ts) discussed performing commanded shocks to test the device¿s ability to deliver a shock successfully.The cause was not conclusively determined and no actions taken at this time.The system remains in service.No adverse patient effects were reported.
|