Event Description
Fill volume: 400.0 ml.Flow rate: 8.0 ml/hr.Procedure: duodenopancreatectomy.Cathplace: epidural.A report was received from an international dist, originating from a hosp, and described that a (b)(6) year old female pt experienced a fast flow and passed away 24 hours after using the product."the pump was installed in the pt and the flow was programmed for 8ml per hour with ropivacaine 0.2%, after 12 hours all the content of 400ml had been infused.The pump has a pca device that was no unlocked.According to the report, the pt had significant hypotension and there is 24 hours that the pt came to death.The volume that should be infused in 50 hours was infused in 12 hours." additional info received (b)(6) 2014: hosp contact confirmed that the cause of death is not due to the use of the device.The pt was ill and had pancreatic cancer.Autopsy was not performed.{2026095-2014-00086/14-00466}.The hosp confirmed that "the pt has a pancreas tumor and developed sepsis after a duodenopancreatectomy".
 
Manufacturer Narrative
Method: the suspect device is being returned for investigation and analysis, but at the time of this report has not been received by the mfr.A review of the device history record (dhr) was conducted for the lot number reported.A photographic image was received from the dist of the suspect unit.Results: results will be provided once the evaluation and investigation have been completed.The dhr review concluded that the lot passed all mfg specifications at release.The photographic image received shows that the red tab of the device is in place.Conclusions: at this time i-flow is pending the return of the device for an evaluation and analysis.The investigation is currently ongoing, once completed a follow-up report will be submitted.The directions for use provide a warning regarding the red tab as indicated below: warning: do not pull the red tab upwards as breakage could occur (figure 6c).If red tab is not removed or breaks while removing, continuous delivery will occur.This delivery may be significantly greater than the total flow rate (bolus+basal).Info from this incident will be included in our product complaint and mdr trend reporting sys.Additional investigation may arise from ongoing analysis, trend info, or other analysis as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RA PCA & SAF 2-14:400 ML, 2-14 ML/HR+5 ML/30 MIN
Type of Device
ELASTOMERIC PUMP
Manufacturer (Section D)
I-FLOW
llc
irvine CA
Manufacturer Contact
maria wagner
43 discovery, suite 100
irvine, CA 92618
9499232324
MDR Report Key3918816
MDR Text Key4514162
Report Number2026095-2014-00086
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K063530
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Distributor
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 06/03/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/28/2014
Device Model Number4440008
Device Catalogue Number101353100
Device Lot Number0200912257
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/03/2014
Initial Date FDA Received06/13/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DRUG: ROPIVACAINE 0.2%
Patient Outcome(s) Death;
Patient Age54 YR
Patient Weight70
-
-