SYNTHES OBERDORF BATTERY OSCILLATOR; BATTERY, REPLACEMENT, RECHARGABLE Back to Search Results
Catalog Number 530.610
Device Problems Device Emits Odor (1425); Smoking (1585); Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/24/2014
Event Type  malfunction  
Event Description
It was reported that prior to surgery, it was observed that the battery oscillator device was heating up and started smoking.The reporter stated that it smelled like an electric fire.There were no delays to the planned surgical procedure as a spare device was available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Additional narrative: the actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the wires on the electronic control unit (ecu) were corroded showing signs of damage.Therefore, the reported condition was confirmed.The assignable root cause was determined to be corrosion due to immersion during cleaning, which is failure to follow the directions for use and would be misuse / abuse and possibly user error.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY OSCILLATOR
Type of Device
BATTERY, REPLACEMENT, RECHARGABLE
Manufacturer (Section D)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH44 36
SZ  CH4436
Manufacturer (Section G)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH44 36
SZ   CH4436
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key3923046
MDR Text Key4517247
Report Number8030965-2014-10373
Device Sequence Number1
Product Code MOQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/24/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.610
Device Lot Number3527
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/03/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/21/2014
Initial Date FDA Received07/09/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/25/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/24/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-