• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALERE SAN DIEGO, INC. INRATIO PT/INR TEST STRIPS; PROTHROMBIN TIME TEST

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALERE SAN DIEGO, INC. INRATIO PT/INR TEST STRIPS; PROTHROMBIN TIME TEST Back to Search Results
Model Number 100139
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Information (3190)
Event Date 05/19/2014
Event Type  Injury  
Event Description
Caller alleged discrepant inratio results.Results as follows: date: (b)(6) 2014, inratio: >7.5, re-test: 4.6, lab: 9.5.All test performed within 5 minutes.The two inratio test results were separate finger sticks performed back to back.Therapeutic range: unk.Coumadin dose was held based on lab result.
 
Manufacturer Narrative
Investigation pending.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INRATIO PT/INR TEST STRIPS
Type of Device
PROTHROMBIN TIME TEST
Manufacturer (Section D)
ALERE SAN DIEGO, INC.
san diego CA
Manufacturer Contact
ya-ling king
9975 summers ridge rd
san diego, CA 92121
8588052084
MDR Report Key3925393
MDR Text Key18932867
Report Number2027969-2014-00578
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092987
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/20/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number100139
Device Lot Number340875
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/20/2014
Initial Date FDA Received06/16/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
INRATIO MONITOR: SERIAL NUMBER (B)(4); JANTOVEN; HYDROCODONE
Patient Outcome(s) Required Intervention;
-
-