This device is used for therapeutic purposes.(b)(4).Product evaluation: analysis found that there was evidence of biological contaminants [based off of the reactivity with hydrogen peroxide].When compared to the assembly drawing: the tip of the outer blade broke off at the first coil of the spiral wrap which would have resulted in the reported malfunction.The portion that became detached measured 0.45¿.The inner blade would spin and middle blade would index freely by hand.When viewed under magnification, there was damage to the hub that is consistent with improper loading; dimples on the front hub prior to the locking area caused by the handpiece locking mechanism.When viewed under magnification, there was deformation of the locking area on the front hub which indicates that at some point the blade was loaded properly and that excess pressure was applied to the blade.It was reported there was some clogging and the blade was repeatedly cleaned with a stylette which reportedly was the likely cause for the failure.There was some minor damage to the tips of the inner and outer blade tips which was likely caused by the repeated cleaning's.Method: microscopic inspection.Results: fracture problem.
|