Facility informed us (service provider) that the collimator came loose and fell striking a pt in the leg while on the x-ray table.We repaired collimator for further use.Condition of pt is unk.We (service provider), were called in to come fix the issue when we were told about the collimator striking a pt.We were not the current service provider of the medical x-ray equipment.Our service engineer remembered the recall from 2007 and repaired according to the recall instructions.Original recall dated 8/3/2007.This report provided by: (b)(6).(b)(4).
|