Model Number ADVIA 1800
Device Problem Low Test Results (2458)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/18/2014
Event Type  malfunction  
Event Description
Discordant, falsely low chloride results were obtained on multiple patient samples on an advia 1800 instrument.The discordant chloride results were reported to the physician(s).The samples were repeated on an alternate system and reported to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant chloride results.
 
Manufacturer Narrative
A siemens customer service engineer (cse) was dispatched to the customer site.The cse evaluated the instrument and instrument data.After instrument evaluation, the cse determined that the equilibration chamber motor was not operational.The cse replaced the equilibration chamber, buffer pump seals and ion selective electrode tubing.The cse calibrated the ion selective electrodes and determined that the customer previously replaced the chloride electrode.The cause of the discordant chloride results was due to a equilibration chamber malfunction.The cse successfully ran quality controls.The instrument is performing according to specifications.No further evaluation of the device is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA 1800
Type of Device
CHEMISTRY ANALYZER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
JEOL LTD, 3-1-2 MUSASHINO AKISHIMA
registration number:3003637681
tokyo, 196- 8558
JA   196-8558
Manufacturer Contact
john nelson
511 benedict avenue
tarrytown, NY 10591
9145242530
MDR Report Key3930458
MDR Text Key4583625
Report Number2432235-2014-00416
Device Sequence Number1
Product Code CGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K990346
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/18/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberADVIA 1800
Device Catalogue Number073-A021-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/18/2014
Initial Date FDA Received07/11/2014
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-