Catalog Number 2C8750
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/20/2014
Event Type  malfunction  
Event Description
It was reported that the second chamber of a clearlink system y-type blood/solution set filled with air and would not drip.The reporter stated that the top chamber was full while the bottom chamber ¿was just sucking air.¿ this occurred during a blood transfusion, when there was approximately 15 minutes left in the procedure.The second lead of the set was not back primed and no saline bag was hung.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
(b)(4).(b)(6).The device is reported to be unavailable for evaluation; however, a photograph of the device was received and will be evaluated.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
(b)(4).Evaluation summary: the physical sample was not returned; however, a photograph was provided for evaluation.During photographic inspection, air was identified in the filter and tubing of the device.A review of the directions for use section of the labeling was performed.The cause of the air in the device was unable to be determined.A capa has been opened to further investigate and address this issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
(b)(4).The cause of the reported no flow was determined to be a use error.It was reported that the unused second lead of the device was not primed.According to the device labeling the user is instructed to prime an unused lead.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCESS
Type of Device
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE - HAINA, SAN CRISTOBAL
parque industrial itabo, zona
haina
DR 
Manufacturer (Section G)
BAXTER HEALTHCARE - HAINA, SAN CRISTOBAL
parque industrial itabo, zona
haina
DR  
Manufacturer Contact
christina arnt
25212 w. illinois route 120
round lake, IL 60073
2242703198
MDR Report Key3935523
MDR Text Key4572833
Report Number1416980-2014-22647
Device Sequence Number1
Product Code BRZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K993120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 06/20/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C8750
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/17/2015
Initial Date FDA Received07/15/2014
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received12/12/2014
02/25/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-