• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET UK LTD. OXF UNI TIB TRAY SZ B LM PMA; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. OXF UNI TIB TRAY SZ B LM PMA; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problem Malposition of Device (2616)
Patient Problem No Code Available (3191)
Event Date 06/12/2014
Event Type  Injury  
Event Description
It was reported patient underwent a left partial knee arthroplasty on (b)(6), 2014.Subsequently, a revision procedure was performed on (b)(6), 2014 due to malpositioning.The tibial bearing and tibial tray were removed and replaced.
 
Manufacturer Narrative
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show that lot released with no recorded anomaly.There are warnings in the package insert that state that this type of event can occur: under warnings, number 1 states, "improper selection, placement, positioning, alignment and fixation of the implant components may result in unusual stress conditions which may lead to subsequent reduction in the service life of the prosthetic components.".
 
Manufacturer Narrative
This follow-up report is being filed to correct information.Discarded.
 
Manufacturer Narrative
This follow-up report is being filed to correct (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF UNI TIB TRAY SZ B LM PMA
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend, south wales IN CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
megan haas
56 e. bell drive
warsaw, IN 46582
5743726700
MDR Report Key3936178
MDR Text Key17618315
Report Number0001825034-2014-06192
Device Sequence Number1
Product Code NRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PP010014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative,company representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 03/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date01/31/2024
Device Model NumberN/A
Device Catalogue Number154720
Device Lot Number857140
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/19/2014
Initial Date FDA Received07/15/2014
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received03/10/2016
03/16/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/28/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-