The customer observed a single non-reproducible, higher than expected vitros ckmb result from a patient sample processed on a vitros 5600 integrated system.A vitros ckmb result of 25.6 ng/ml vs.Expected result of 0.54 ng/ml was obtained from the sample.A biased result of the direction and magnitude observed may lead to inappropriate physician action if undetected.The non-reproducible, higher than expected ckmb result was reported from the laboratory, however, a corrected report was issued and no treatment was given, changed, or withheld based on the reported ckmb result reported.There was no allegation of patient harm.(b)(4).
|
The investigation confirmed that a non-reproducible, higher than expected vitros ckmb result was obtained from a patient sample processed on the vitros 5600 system.The investigation could not determine a definitive root cause.There was no indication the vitros 5600 system or the vitros ckmb reagent contributed to the event.Additionally, it is unknown if the sample was processed in accordance with the tube manufacturer¿s recommendations.It is possible that cellular debris, due to poor sample preparation, was present in the affected sample, although this could not be confirmed.A definitive root cause for the event could not be determined.
|