• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS, INC. MERIT CUSTOM KIT; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS, INC. MERIT CUSTOM KIT; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Catalog Number K09-MST2176A
Device Problems Delivered as Unsterile Product (1421); Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/03/2014
Event Type  malfunction  
Event Description
The distributor reported there is a small hole in the tyvek of the package causing a breach in the sterile barrier.This was identified during their initial inspection of received product.The device was not sent to a suer facility.
 
Manufacturer Narrative
Device evaluation: one new device was returned for evaluation.The device history record was reviewed and no exception documents were found.The complaint database was reviewed and no similar complaints for this lot number were found.The returned device was examined and the complaint is confirmed.A hole was found in the tyvex.The tyvek appears to be thin and the extra stress on the package resulted in a tear that was not part of the seal.The hole appears to have been caused by rough or improper handling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERIT CUSTOM KIT
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS, INC.
south jordan UT
Manufacturer Contact
jerry mcphie
1600 west merit pkwy.
south jordan, UT 84095
8012084491
MDR Report Key3944074
MDR Text Key4535888
Report Number1721504-2014-00089
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
PMA/PMN Number
K913682
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative,Distributor
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 04/03/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2017
Device Catalogue NumberK09-MST2176A
Device Lot NumberH592977
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer04/16/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/03/2014
Initial Date FDA Received05/01/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-