• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: J. T. POSEY CO. KEEPSAFE DELUXE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

J. T. POSEY CO. KEEPSAFE DELUXE Back to Search Results
Model Number 8374
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
The customer reported that the connection for the sensor on the alarm is working occasionally.The customer did not know when discovered or the date when found.No pt incident or injury was reported.
 
Manufacturer Narrative
Evaluation codes : results - evaluation of the returned product confirmed the reported issue; unit sounds on and off when the sensor cable is wiggled.Unit passe all other functional testing.All pins inside the sensor receptacle are bent down.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KEEPSAFE DELUXE
Manufacturer (Section D)
J. T. POSEY CO.
arcadia CA
Manufacturer (Section G)
J. T. POSEY CO.
5635 peck rd.
arcadia CA 91006
Manufacturer Contact
pam wampler, admin
5635 peck rd.
arcadia, CA 91006
0626443314
MDR Report Key3952959
MDR Text Key21875498
Report Number2020362-2014-00193
Device Sequence Number1
Product Code KMI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Unknown
Type of Report Initial
Report Date 04/22/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8374
Device Catalogue Number8374
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer05/09/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/22/2014
Initial Date FDA Received05/20/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-