(b)(4).Method: the complaint mr290v vented humidification chamber was returned to fisher & paykel healthcare in (b)(4) where it was visually inspected.Results: visual inspection revealed that the feedset tube was separated from the chamber at the connection to the chamber dome.Glue was noted around the feedset tube.A lot check revealed no other complaints of this nature for lot 041119.Conclusion: we are unable to determine what may have caused the feedset tube to separate from the chamber dome.All chambers are pressure tested before they leave the production line and any holes or leaks in the feedset are identified during this process.Any chamber that fails is rejected.This suggests that the feedset tube separated from the chamber dome after it was released for distribution.We have conducted extensive testing of the mr290 chamber, with particular emphasis on feedset breaks.Significantly we have not been able to replicate failure of the feedset tube at the chamber dome in any of our testing.The specification for the chamber requires that the feedset tube should have a breaking strain of 30 newtons.During production, pull testing of the feedset strength at both spike and dome end is performed every hour on mr290 chambers from each production line.Any chamber that fails is rejected and the whole batch is placed on hold for investigation.The user instructions that accompany the mr290 state the following: -"set appropriate ventilator alarms." -"perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.".
|