It was reported that the event occurred during use while in the cathlab.The md inserted the iab-06840-u through the teflon sheath via right femoral artery and felt a little resistance during the insertion.Several minutes after pumping was started, light leak/helium loss 2 was displayed on the screen.The user checked the gas, but no problem was found.As a result, the iab was removed successfully.A new kit was opened and inserted through the same insertion site (right femoral artery) without issue and intra-aortic balloon pump (iabp) therapy continued as planned.Pump strips were generated.The iabp used during this event was an iabp-0500j, serial number (b)(4).There was no report of patient death, complications or injury.No medical/surgical intervention was required.It was stated there was no delay or interruption in therapy; however, there was the time to remove, prep and insert a new iab.The patient outcome is good.
|