• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GE OEC MEDICAL SYSTEMS (SLC) 9900; FLUOROSCOPIC X-RAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE OEC MEDICAL SYSTEMS (SLC) 9900; FLUOROSCOPIC X-RAY Back to Search Results
Model Number 9900
Device Problems Device Displays Incorrect Message (2591); Reset Problem (3019)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/29/2014
Event Type  malfunction  
Event Description
The customer reported the system displayed a software corruption error message.The field engineer reported that this error resulted in the system not being able to boot completely.There is no report of injury or death associated with the event described in this complaint.
 
Manufacturer Narrative
A ge service representative performed an onsite investigation.The system hard drive was evaluated, replaced and reformatted.The system software and calibration files were reloaded.The system was tested and found to be working as intended and returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9900
Type of Device
FLUOROSCOPIC X-RAY
Manufacturer (Section D)
GE OEC MEDICAL SYSTEMS (SLC)
384 wright brothers dr.
salt lake city UT 84116
Manufacturer (Section G)
GE OEC MEDICAL SYSTEMS(SLC)
384 wright brothers dr.
salt lake city UT 84116
Manufacturer Contact
384 wright brothers dr.
salt lake city, UT 84116
8015364952
MDR Report Key3960274
MDR Text Key4606373
Report Number1720753-2014-05380
Device Sequence Number1
Product Code JAA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/23/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9900
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/16/2014
Initial Date FDA Received06/24/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-