The product in question was not returned for evaluation, therefore an inspection to ensure the witness lines of the crimped stent were visible was not able to be performed (when the product is crimped, the stent struts impart permanent imprints to the exterior of the balloon).The diameter of the stent could not be measured to ensure that the diameter matched the measurements taken during the quality final inspection.The introducer sheath used in the case was not returned as well thus an inspection of the sheath dimension and overall condition could not be assessed.A full of review of the catheter lot history records for the device in question was performed.The records indicate that this lot of catheters passed atriums final lot qualification testing.This inspection requires that the catheter lot must pass the following.Ability of the stent and delivery system to be passed through the labeled introducer sheath.Ability to deploy the stent and nominal pressure (8atm).Ability to withdraw the deflated balloon catheter back through the labeled introducer sheath.Ability of the delivery system to withstand 5 inflate.Result: the details of the complaint indicate that the stent was dislodged while attempting to pass the stent through a tight lesion.It is possible that the lesion was of a diameter that was smaller than the crimped stent outer diameter of the i-cast covered stent.This could cause the stent to dislodge if enough force was generated while attempting to pass the stent through the lesion.
|